Skip to main content
. 2020 Jan 17;17:8. doi: 10.1186/s12986-020-0427-7

Table 3.

Assessment of the bias of 25(OH)D2 detected by LC-MS/MS using NIST certified SRM-972a

SRM-972a
25(OH)D2
Certified values by NIST LC-MS/MS
Target values (ng/mL) Uncertainty Detected values (ng/mL) SD (n = 3) Bias(%)
Level 1 0.50 0.06 0.48 0.07 −4.17
Level 2 0.81 0.06 0.84 0.05 4.45
Level 3 12.90 0.30 12.72 0.26 −1.40
Level 4 0.50 0.10 0.51 0.04 2.26

SRM standard reference material, NIST National Institute of Standards & Technology