Skip to main content
. 2020 Jan 21;10:800. doi: 10.1038/s41598-020-57680-w

Table 2.

Cross-classification of Regions Of Interest (ROIs) identified as cysts or Non-Perfused Areas (NPAs) with structural B scan with flow as compared with the presence or absence of Suspended Scattering Particles in Motion (SSPiM) with OCT Angiography using paired ROIs identified using both PLEX ELITE 9000 and ANGIOVUE.

SCP total % cyst
cyst NPA
PLEX ELITE 9000 SSPiM 13 3 16 81%
no 0 21 21 0%
ANGIOVUE SSPiM 5 4 9 56%
no 5 23 28 18%
DCP
cyst NPA total % cyst
PLEX ELITE 9000 SSPiM 14 3 17 82%
no 0 20 20 0%
ANGIOVUE SSPiM 5 3 8 63%
no 0 29 29 0%