Table 3.

Repeatability (intra‐assay variability) of three serum samples from dogs analyzed on four different analyzers performed 10 times in one run. None of the variabilities was less than 10%, which is a general target value for repeatability, and only three of the 12 variabilities were within the "poor but acceptable" range of 10%‐20%

Sample# Spec cPL Analyzer Measurements %CV
µg/L
1 120 1 10 22.9
2 10 19.4
3 10 31.5
4 10 23.0
2 233 1 10 26.0
2 10 34.7
3 10 30.3
4 10 22.3
3 588 1 10 18.7
2 10 34.6
3 10 21.0
4 10 16.9
Mean       25.1
SD       6.2