. 2020 Jan 21;15(1):e0227694. doi: 10.1371/journal.pone.0227694

Table 2. The concordance of urine albumin/creatinine ratio using the semi-quantitative method and quantitative testing.

Quantitative testing
(mg/g)
semi-quantitative method Sensitivity Specificity *Over detected fraction **Under detected fraction
<30 30–300 ≥300
All diabetes
(n = 1,881)
<30 664 (61.4) 408 (37.7) 9 (0.8) 93.5 61.4 22.2 2.8
30–300 52 (10.8) 350 (72.9) 78 (16.2)
≥300 0 (0.0) 17 (5.3) 303 (94.7)
Diabetes with eGFR ≥60 ml/min/m2 and dipstick (-)
(n = 1,110)
<30 579 (62.9) 333 (36.3) 5 (0.5) 81.3 63.1 30.5 3.2
30–300 36 (19.4) 142 (76.3) 8 (4.3)
≥300 0 (0.0) 2 (28.6) 5 (71.4)

*, number of false positives/total number of population x 100

**, number of false negatives/total number of population x 100

eGFR, estimated glomerular filtration rate