Skip to main content
. 2019 Nov 14;34(1):83–91. doi: 10.1111/jvim.15657

Table 3.

Results of Freestyle's analytical accuracy in the low (BG < 100 mg/dL) and high glucose range (BG ≥ 100 mg/dL) before and after DKA resolution

Before DKA resolution After DKA resolution
Low glucose range (BG <100 mg/dL)
n 5 21
MAD (mg/dL) 34.0 22.7
Percent of values within ±15 mg/dL of the BG value 0% (0/5) 42.9% (9/21)
High glucose range (BG ≥100 mg/dL)
n 224 235
MARD (%) 19.7 17.2
mARD (%) 18.2 13.7
MRD (%) −3.5 −6.0
Percent of values within ±15% of the BG value 43.8% (98/224) 53.6% (126/235)

Abbreviations: BG, blood glucose; DKA, diabetic ketoacidosis; MAD, mean absolute difference; MARD, mean absolute relative difference; mARD, median absolute relative difference; MRD, mean relative difference.