Skip to main content
. 2019 Dec 19;17(1):54. doi: 10.3390/ijerph17010054

Table 3.

Correlation and relationship significance analysis between input and output parameters of the parametric study.

Variable A1 Time
(ms)
A1 Length
mm)
A1 Velocity
(mm/s)
HC Time
(ms)
Peak Dist.
(mm)
A1 Deformation Amp. (mm) HC Deformation Amp. (mm) AP1 (mmHg) SP-HC Stiffness parameter
IOP (mmHg) Pearson Correlation (r) 0.725 ** −0.455 ** −0.731 ** −0.255 ** −0.616 ** −0.403 ** −0.635 ** 0.736 ** 0.442 **
Sig. (two-tailed) 0.000 0.000 0.000 0.007 0.000 0.000 0.000 0.000 0.000
CCT (µm) Pearson Correlation (r) 0.382 ** 0.637 ** −0.206 * −0.122 −0.500 ** 0.673 ** −0.493 ** 0.385 ** 0.468 **
Sig. (two-tailed) 0.000 0.000 0.031 0.204 0.000 0.000 0.000 0.000 0.000
μ Pearson Correlation (r) 0.338 ** 0.471 ** −0.367 ** −0.280 ** −0.407 ** 0.432 ** −0.377 ** 0.355 ** 0.434 **
Sig. (two-tailed) 0.000 0.000 0.000 0.003 0.000 0.000 0.000 0.000 0.000
R (mm) Pearson Correlation (r) −0.007 −0.056 −0.067 0.032 0.088 −0.253 ** −0.052 0.007 −0.088
Sig. (two-tailed) 0.946 0.564 0.486 0.741 0.362 0.008 0.592 0.945 0.362

Note: IOP, intraocular pressure; CCT, central corneal thickness; μ, corneal material stiffness coefficient; R, corneal curvature radius; A1 is the first applanation; HC, highest concavity; AP1, first applanation pressure; SP-HC, stiffness parameter at highest concavity. **: Correlation is significant at the 0.01 level. *: Correlation is significant at the 0.05 level.