Skip to main content
. 2020 Jan 28;58(2):e01478-19. doi: 10.1128/JCM.01478-19

TABLE 2.

Comparison of PCR CT values for M. genitalium detection when the IC of the evaluated kit was added to the specimen (before DNA extraction) alone versus in addition to the other two ICsa

Serial dilution of M. genitalium G37 culture (genome equivalent/μl) Mean CT ± SD
S-DiaMGRes (Diagenode)
RealAccurate TVMGres (Pathofinder)
ResistancePlus MG (SpeeDx)
Diagenode IC alone Three ICs ΔCT Pathofinder IC alone Three ICs ΔCT SpeeDx IC alone Three ICs ΔCT
7.2 × 104 20.80 ± 0.21 21.43 ± 0.04 +0.63 20.71 ± 0.32 21.73 ± 0.39 +1.02 13.01 ± 0.00 12.96 ± 0.01 −0.05
1.2 × 103 26.48 ± 0.26 25.86 ± 0.20 −0.62 24.64 ± 0.40 24.81 ± 0.85 +0.17 17.02 ± 0.01 16.54 ± 0.14 −0.48
3.0 × 102 28.94 ± 0.52 28.00 ± 0.27 −0.94 27.29 ± 0.12 27.83 ± 0.17 +0.54 22.04 ± 0.02 21.98 ± 0.02 −0.07
2.8 × 101 34.38 ± 0.35 35.16 ± 0.55 +0.78 33.75 ± 0.04 34.44 ± 0.30 +0.69 26.47 ± 0.42 26.89 ± 0.54 +0.42
a

Experiments were performed in triplicate.