Table 2. MRI sequence parameters.

Field Strength Sequence FOV [mm2] Resolution [mm3] TR/TE/TI [ms] Duration [min]
1.5 T T2w STIR 340 × 340 0.8 × 0.8 × 4 4900/62/165 2.33
Dynamic T1w GRE (6×) 360 × 360 0.8 × 0.8 × 1.5 7.7/4.77 1.07×6
DWI SE-EPI SPAIR 340 × 170 1.8 × 1.8 × 4 5100/60/150 3.29
3.0 T T2w STIR 340 × 340 0.8 × 0.8 × 4.0 3570/70/230 3.29
Dynamic T1w GRE (6×) 360 × 360 0.8 × 0.8 × 1.5 5.97/2.46 1.03×6
DWI SE-EPI SPAIR 350 × 185 1.8 × 1.8 × 2.5 4300/58 1.53

The protocol was optimized following international recommendations and current practice in breast MRI [5,13]. Images were acquired in axial plane. B-values were: 50, 400, and 800 s/mm2. FOV: Field of view. TR, TE, TI: Repetition, echo, inversion time. T: Tesla. STIR: Short Tau Inversion Recovery. GRE: Gradient echo. SE: Spin echo. EPI: Echo planar imaging. SPAIR: Spectral adiabatic inversion recovery.