. 2020 Jan 22;61(2):129–136. doi: 10.3349/ymj.2020.61.2.129

Table 4. Diagnostic Properties Regarding the WHUS and Standard Devices.

AUC Cutoff value (mm) Sensitivity Specificity PPV NPV Estimated number of CAD patients using cutoff Number of true CAD
max-CIMT of standard device 0.807 1.05 0.78 0.53 0.98 0.29 60 39 (97.5%)
max-CIMT of WHUS 0.725 1.10 0.98 0.27 0.78 0.53 63 39 (97.5%)

max-CIMT, mean of maximal value of carotid intima-medial thickness of a patient; AUC, area under the curve; PPV, positive predictive value; NPV, negative predictive value; WHUS, wireless handheld ultrasound; CAD, coronary artery disease.