Table 2.

Comparison of blood pressure measurement using TOF-Cuff® and standard monitoring (MP 30, Philips; Zurich, Switzerland)

Time point of assessment TOF-Cuff® Standard monitoring P value
Beginning of anesthesia induction Systolic 143.0 (130.5–151.0) 133.0 (117.8–149.3) 0.1781
MAP 94.0 (86.0–104.0) 89.5 (81.8–97.3) 0.0791
Diastolic 63.0 (58.5–70.0) 74.5 (68.8–85.0)  < 0.0011
End of anesthesia induction Systolic 107.0 (89.8–117.3) 109.0 (93.8–117.0) 0.8051
MAP 72.0 (61.8–84.0) 76.5 (69.8–80.8) 0.9411
Diastolic 51.5 (41.8–57.3) 64.0 (54.5–73.5)  < 0.0011
All Systolic 124.6 (± 29.8) 123.0 (± 29.4) 0.6162
MAP 85.5 (± 20.3) 84.2 (± 16.6) 0.4882
diastolic 56.2 (± 13.5) 70.7 (± 15.5)  < 0.0012

Data displayed as mean (± SD) or median (IQR) in mmHg; all assessments for blood pressure that were based on two measurements are weighted by weight = 0.5

MAP mean arterial pressure

1Exact Wilcoxon signed-rank test

2One-sample weighted t test