Table 2.

Absolute dose at different field dimensions comparing with dose calculated by treatment planning system PRECISE ELEKTA.

Energy factor Field size Calc. dose/100MU Measured (ROOS IC) (mean ± SD)
6 MeV 6 × 6 78.9 85
8 × 8 99.7 99
10 × 10 94.4 102
14 × 14 94.4 100 ± 0.3
6 × 14 94.4 100 ± 0.5
8 × 14 95.62 102
6 × 17 95.62 100 ± 0.2
8 × 17 95.62 100 ± 0.3
8 MeV 6 × 6 86.3 85.5
8 × 8 99.4 100
10 × 10 96.3 99.7
14 × 14 96.3 99 ± 0.5
6 × 14 96.3 100.6 ± 0.5
8 × 14 96.5 100
6 × 17 96.5 98 ± 0
8 × 17 96.5 100 ± 0
10 MeV 6 × 6 84.1 93
8 × 8 99.9 100
10 × 10 94.9 99
14 × 14 94.9 96
6 × 14 94.9 99
8 × 14 94.2 98
6 × 17 94.2 96
8 × 17 94.2 99
15 MeV 6 × 6 95.6 98
8 × 8 99.8 100
10 × 10 95.4 99
14 × 14 95.4 98
6 × 14 95.4 100
8 × 14 94.7 98
6 × 17 94.7 99
8 × 17 94.7 100