. 2019 Nov 18;58(2):239–247. doi: 10.1007/s11517-019-02067-x

Table 2.

Data on drift of tcPO2 (24-h calibration interval) and tcPCO2 (12-h calibration interval)

Total drift during calibration interval (12 h/24 h) (%) Drift during first hour of calibration interval (%/h) Drift during last hour of calibration interval (%/h)
tcPO2 0.30 (0.63) 0.14 (0.28) 0.03 (0.21)
tcPCO2 − 2.53 (1.04) 0.49 (0.28) 0.18 (0.09)

Data is shown as mean (SD)