Table 2.
Parameter | Acceptance criteria | Recommendations/ comments |
---|---|---|
PCR Efficiency | 90%‐110% |
Absolute quantification: 100 ± 10% Semi‐quantification (eg, comparative quantification of extracellular miR level): ±10% of the mean of batches to be compared; mean does not necessarily have to be 100% |
Slope of standard curve | −3.1 to −3.6 | Slope up to −4.1 is acceptable, when it is constant within batches to be compared |
Correlation coefficient (R2) | ≥ 0.98 | Correlation coefficient lower than 1 result in inaccuracy quantification of at least one calibration standard |
Linearity | see R 2 | Within calibration range of extracellular miR concentrations; Aa [%] of calculation standards ± 30%, but can be higher when R 2 < 1.0 |
Specificity | Tmb ± 1°C | Melting curve analysis of all batches |
Sensitivity | ALLOQ c ≤ 30% | Accuracy of standard curve can be higher when R2 < 1.0 |
Repeatability |
CVd ≤ 25% CVLLOQ ≤ 30% |
CV of quantified concentrations, not Cq‐values |
Reproducibility |
CV ≤ 35% CVLLOQ ≤ 40% |
CV of quantified concentrations, not Cq‐values |
Trueness | AQC ± 30% | Accuracy of low QCse is not within acceptance criteria |
Accuracy.
Melting temperature.
Lower limit of quantification.
Coefficient of variation.
Quality control.