Table 3.

Overall test characteristics

Test variable Anti-PLA2R ChLIAa Anti-PLA2R ELISAa Anti-PLA2R RC-IFAa
Area under the curve (95% CI) 0.899 (0.854–0.944) 0.927 (0.894–0.961) 0.916 (0.880–0.952)
Maximum sum of sensitivity and specificity, % (cutoff) 183.9 (9.1) 181.3 (3.0) 183.2 (1:5)
Sensitivity at 98% specificity, % (cutoff) 85.2 (7.8) 81.9 (5.8) NA
Sensitivity at 99% specificity, % (cutoff) 84.5 (9.1) 80.7 (7.3) NA
Sensitivity at 100% specificity, % (cutoff) 83.2 (15.5) 76.1 (15.2) 83.2 (1:5)

Anti-PLRA2, anti–phospholipase A2 receptor; ChLIA, chemiluminescence immunoassay; CI, confidence interval; ELISA, enzyme-linked immunosorbent assay; NA, not available; RC-IFA, recombinant cell-based indirect immunofluorescence assay; RU, relative units.

a

Cutoff values are indicated in chemiluminescent units per milliliter (ChLIA), RU/ml (ELISA), and titer (RC-IFA).