Skip to main content
. 2019 Nov 13;47(1):223–233. doi: 10.1002/mp.13883

Table 4.

Differences between MC and experimental dosimetric parameters for EBT3 films irradiations parallel to the proton beam for magnetic fields strengths of B = 0 and 1 T.

Beam B (T) ΔR80 (mm) ΔDmean/ ΔDmax (%) Δμmean (mm) ΔFWTMmea0 (mm)
Single beams
E = 97.4 MeV 0 −0.1 0.5/6.4 0.2 0.8
1 0.1 −0.2/7.7 −0.4 −0.7
E = 148.2 MeV 0 −0.1 −0.1/5.6 −0.3 0.8
1 −0.1 −0.9/5.9 −0.7 0.7
E = 198 MeV 0 0.1 −0.6/4.9 −0.1 −0.3
1 −0.1 −0.5/8.2 −0.9 −0.4
SOBP 40 × 40 × 40 mm3
Centered at 75 mm 0 −0.1 0.3/5.9 −0.1 −0.1
1 −0.1 −0.5/7.3 0.4 0.2
Centered at 115 mm 0 0.1 0.1/7.2 0.4 −0.8
1 0.1 −0.5/8.7 0.8 −1.0