Skip to main content
. 2019 Aug 28;5(1):62–69. doi: 10.1016/j.adro.2019.08.005

Table 2.

Mean percent change in dosimetric parameters of planning target volume coverage for the 3 TTFields array positions relative to coverage without the array in place

Case V97% V96% V95% D97% D96% D95%
1 −0.9% −0.6% −0.4% −1.5% −1.4% −1.3%
2 −0.1% −0.1% 0.0% −0.6% −0.6% −0.6%
3 0.0% 0.0% 0.0% −0.9% −0.9% −0.9%
4 0.3% 0.2% 0.2% 0.0% −0.1% −0.1%
5 −1.7% −0.4% −0.1% −2.7% −2.7% −2.7%
6 0.1% 0.0% 0.0% 0.3% 0.4% 0.4%
7 −0.6% −0.2% −0.1% −1.3% −1.3% −1.3%
8 −0.2% 0.0% 0.0% −1.0% −1.0% −1.0%
9 0.3% 0.2% 0.1% 0.7% 0.6% 0.5%
10 −0.3% −0.1% 0.0% −1.3% −1.3% −1.2%
Mean −0.3% −0.1% 0.0% −0.8% −0.8% −0.8%
SD 0.6% 0.3% 0.1% 1.0% 1.0% 0.9%

Abbreviation: TTFields = tumor-treating fields.