Fig. 3.

Fig. 3

sO2 estimates using 26 wavelengths between 500 and 1000 nm. Top row: xz slice of error in oxygenation midway along tube’s length when the background is spectrally flat and the error originates from spectral coloring of the fluence in the blood-filled tube. Dashed lines used to indicate four xy slices (a–d) from which histograms of sO2 are formed (true value of 90% indicated by black line). Bottom row: xz slice of error in oxygenation midway along tube’s length when the background optical properties vary with wavelength. Both the cases with and without a depth-dependent fluence correction are shown. Dashed lines used to indicate four xy slices (a–d) from which histograms of sO2 are formed (true value of 90% indicated by vertical black line). (a–d) sO2mean represent the mean of the sO2 estimates of the voxels contained within the tube in slices a, b, c, and d, respectively, for the corrected data. The global mean sO2g.mean of the corrected data, which represents sO2 estimates averaged over all voxels enclosed within the entire tube, is shown to the right of the histograms.