Skip to main content
. Author manuscript; available in PMC: 2021 Feb 1.
Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2019 Sep 19;67(2):248–257. doi: 10.1109/TUFFC.2019.2942557

TABLE I.

Imaging Parameters for Volumetric Data Acquisition

Type Parameters Value
Application-specific requirements Volume rate 30 Hz
Volume data acquisition rate 750 Hz
Total no. of acquired volumes 100
Design parameters No. of synthetic transmit/receive events 16
No. of virtual sources 25 (5×5)
Transmit diverging angle 30°
Angle interval between virtual sources 7.5°
Focal distance of a central virtual source −19 mm
Pulse-repetition interval 83.25 μs
Transmit parameters Center frequency 5.2 MHz
No. of cycles of Tx 1 cycle
Applied voltage 20 V
System Resource Required system memory > 12.2 GB