Skip to main content
. 2020 Feb;41(2):231–237. doi: 10.3174/ajnr.A6375

Table 2:

Quantitative analysisa

Non-SEMAR SEMAR P Value
Severity of metallic artifacts
 SD around DBS electrodes (HU) 28.3 ± 9.1 9.8 ± 3.2 <.01
 Artifact index (HU) 28.0 ± 9.2 9.0 ± 3.5 <.01
Background image quality
 GM-WM contrast (HU) 11.7 ± 2.7 11.6 ± 2.4 .35
 Image noise (HU) 3.8 ± 0.7 3.7 ± 0.7 .17
 CNR 3.2 ± 1.0 3.2 ± 0.8 .65

Note:—CNR indicates contrast-to-noise ratio; HU, Housfield unit.

a

Data are means.