Table VII.

Summary of noise level in virtual single-energy computed tomography (vSECT) images at default and optimized dose partitioning factor (α) for biphasic liver imaging; note: the percentages in the parentheses indicate the noise increase (+) or decrease (−) on vSECT images with optimized αI,Gd in reference to the default α.

Tube potential pair (kV) Biphasic liver imaging
Noise at default α
Noise at optimized αI,Gd
V120kV,I V120kV,Gd V120kV,I V120kV,Gd
80/Sn150 67.6 79.7 61.9 (−8.4%) 69.3 (−13.0%)
90/Sn150 99.6 110.3 88.8 (−10.8%) 94.2 (−14.6%)
100/Sn150 153.5 164.8 118.6 (−22.7%) 121.8 (−26.1%)