Published in final edited form as: Med Phys. 2019 Dec 10;47(2):441–456. doi: 10.1002/mp.13912

Table VIII.

Summary of noise level in virtual single-energy computed tomography (vSECT) images at default and optimized dose partitioning factor (α) for small bowel imaging; note: the percentages in the parentheses indicate the noise increase (+) or decrease (−) on vSECT images with optimized αI,Bi in reference to the default α.

Tube potential pair (kV) Small bowel imaging
Noise at default α
Noise at optimized αI,Bi
V120kV,I V120kV,Bi V120kV,I V120kV,Bi
80/Sn150 14.9 19.1 15.3 (+2.7%) 16.8 (−12.0%)
90/Sn150 15.7 19.5 16.5 (+5.1%) 17.5 (−10.3%)
100/Sn150 20.8 25.5 20.0 (−3.8%) 19.5 (−23.5%)