Validation result summary
Test Parameter Acceptance Criteria Result
Accuracy Recovery 90–110% ([18F]FET spiked with [18F]fluoride) [18F]fluoride concentration Recovery
1% 105.0%
3% 104.6%
5% 105.9%
Repeatability 6 repetitions (5% [18F]fluoride) RSD ≤5% RSD = 1.96%
Intermediate Precision RSD ≤5% (3 analysts) RSD = 2.90%
Specificity Resolution of > 2 between peaks Resolution of[18F]fluoride and [18F]FET Rs = 5.7
Limit of Quantification S/N ratio ≥ 10 24 kBq/ml
Linearity 12–1232 kBq/ml (5 concentrations in triplicate) R > 0.99 R = 0.995
Range Reported value 4.8–1232 MBq/ml