Table 3. Comparison of HRP Immobilization Methods on Solid Supports for H2O2 Sensing.

support of HRP technique/reagent LODs (nM), linear range (μM) portable reusable (uses) stability (months) sample refs
PDMS-TEOS-SiO2-TMB spectroscopy/TMB 400, 1.2–72 yes no 3 cosmetics (20)
Cu3(PO4)2-3H2O spectroscopy/TMB 200, 0–50 no no   urine (34)
PEG fluorescence/HP green 9100, 30–300 no yes 1 cell nutrition medium (29)
UVNPs@PSIOAm luminescence 64, 0.1–5 no no   serum (31)
sol–gel luminescence -, 100–3000 no yes (5) 2 contact lenses solutions (33)
PDMS luminescence 20, 0.06–10 yes yes (60) 3 cell culture medium cleaning denture tablet this work
AgNPs/NDs/GCE amperometry 10, 0.1–34 no no 1 commercial eye drops (12)
PGN/GCE-2 amperometrya 0.0267, 0.00008–0.664 no yes (50)   living cells (24)
MIL-100(Cr)-B amperometry 100, 0.5–3000 no yes (100) 1 living cells (26)
nafion-nile blue amperometry 280, 0.5–1.5 no no   no (27)
a

Logarithmic calibration.