1. Patient preparation |
1.1 |
Subject should be scanned in a normal hydration status when clinically appropriate |
24 |
100 |
2. Hardware |
2.1 |
Both 1.5 T and 3 T are adequate field strengths |
0 |
87 |
2.2 |
The body coil should be used as transmitter coil |
6 |
93 |
2.3 |
Body phased-array coils should be used as receive coils |
0 |
94 |
3. Labeling strategy |
3.1 |
Both PASL:FAIR and PCASL are adequate labeling strategies |
6 |
93 |
3.2 |
Single time point acquisitions are recommended for simplicity of acquisition and data analysis |
6 |
86 |
3.3 |
Multiple time point acquisitions require a longer acquisition time and more complicated processing. However, they can provide measurements of perfusion and ATT that can be useful if delayed arrival time is suspected in a clinical population |
6 |
100 |
4. FAIR labeling parameters |
4.1 |
A FOCI pulse should be used for the selective inversion to optimize the inversion slice profile |
19 |
92 |
4.2 |
The selective slab should be carefully positioned, excluding the aorta |
6 |
100 |
4.3 |
The selective inversion slab thickness should equal the imaging slab thickness + [10–20] mm |
13 |
86 |
4.4 |
In single-TI acquisitions, an inversion time of 1.8–2.0 s is recommended |
10 |
89 |
4.5 |
In single-TI acquisitions, an approach for controlling the temporal width of the bolus (QUIPSS II or Q2TIPS) must be used to quantify perfusion |
25 |
100 |
4.6 |
A bolus duration (TI1) of 1.0–1.2 s is recommended |
25 |
92 |
4.7 |
In single-TI acquisitions, a minimum of 20 ASL pairs is recommended |
10 |
89 |
5. PCASL parameters |
5.1 |
An unbalanced version of PCASL is preferred due to its lower sensitivity to off-resonance effects |
38 |
80 |
5.2 |
A labeling time of 1.5–1.8 s is recommended |
10 |
100 |
5.3 |
The labeling plane should be oriented approximately perpendicular to the aorta |
13 |
100 |
5.4 |
The labeling plane should be positioned at approximately 8–10 cm from the centre of the kidney, in the superior direction |
14 |
94 |
5.5 |
Hanning RF pulses are recommended |
31 |
100 |
5.6 |
An RF pulse duration of 500 μs is recommended |
25 |
100 |
5.7 |
Pulse spacing (from the centre of one pulse to the centre of the next) of 1000 μs or shorter is recommended |
31 |
100 |
5.8 |
Average B1 of 1.6 μT is recommended |
27 |
100 |
5.9 |
Average gradient (Gave) of 0.4–0.6 mT/m is recommended |
25 |
92 |
5.10 |
Gmax to Gave Ratio of 6–7 is recommended |
40 |
100 |
5.11 |
In single PLD acquisitions, a PLD of 1.2–1.5 s is recommended |
19 |
100 |
5.12 |
In single-PLD acquisitions, a minimum of 20 ASL pairs is recommended |
14 |
83 |
6. Readout |
6.1 |
A 2D single-slice acquisition scheme is recommended as the default renal ASL method |
10 |
95 |
6.2 |
Multislice 2D acquisition schemes are recommended for applications that require extended kidney coverage |
5 |
100 |
6.3 |
3D acquisition schemes represent a promising alternative to 2D multislice schemes but are not recommended as the default method for extended kidney coverage due to limited clinical experience with 3D schemes |
10 |
95 |
6.4 |
Spin-echo EPI is the preferred readout for 2D single-slice acquisitions |
5 |
75 |
6.5 |
bSSFP and single-shot RARE are adequate alternatives to EPI for 2D single-slice acquisitions |
14 |
94 |
6.6 |
Spin-echo EPI is the preferred readout for 2D multislice acquisitions |
14 |
83 |
6.7 |
Coronal oblique slices (along the major axis of the kidneys) are preferable for renal ASL |
6 |
93 |
6.8 |
The recommended slice thickness in 2D acquisitions is 4-8 mm |
19 |
100 |
6.9 |
The recommended slice thickness in 3D acquisitions isss 3-6 mm |
13 |
100 |
6.10 |
The recommended in-plane resolution is 2-4 mm |
0 |
93 |
6.11 |
Undersampling methods, such as partial Fourier and parallel imaging at moderate acceleration factors (up to R = 2) may be used |
19 |
100 |
6.12 |
The recommended TR (including labeling + readout) is 4–6 s |
0 |
94 |
7. Other sequence details |
7.1 |
Pre and post-inversion saturations are recommended for FAIR labeling schemes |
14 |
100 |
7.2 |
Background-suppression is recommended for renal ASL |
5 |
80 |
7.3 |
Breath-hold scans are not recommended for clinical renal ASL |
0 |
94 |
7.4 |
Renal ASL scans under free breathing are preferred |
0 |
76 |
7.5 |
Respiratory triggering can be advantageous to minimize the effects of kidney motion at the expense of scan time |
5 |
95 |
7.6 |
Fat suppression is recommended for renal ASL |
5 |
90 |
8. Data preprocessing |
8.1 |
Retrospective image registration is highly recommended for renal ASL |
13 |
100 |
8.2 |
Outlier rejection is recommended for renal ASL |
0 |
100 |
9. Quantification |
9.1 |
M0 acquisition is mandatory |
0 |
94 |
9.2 |
Using a single-compartment model with assumed blood T1 for quantification is recommended for robustness and simplicity |
7 |
100 |
9.3 |
A two-compartment model with separate transit time and tissue T1 measurements is a viable alternative to the single-compartment approach but requires more complex acquisition/analysis methods and therefore is currently not recommended as the default renal ASL approach |
10 |
95 |
9.4 |
Tissue-blood partition coefficient = 0.9 mL/g [82–84, 97] |
5 |
90 |
9.5 |
Assumed blood T1 at 3 T = 1.65 s [98] |
0 |
100 |
9.6 |
Assumed blood T1 at 1.5 T = 1.48 s [98] |
13 |
93 |
9.7 |
Labeling efficiency PASL = 95% (neglecting background suppression loss) |
6 |
100 |
9.8 |
Labeling efficiency PCASL = 85% (neglecting background suppression loss) |
13 |
86 |
9.9 |
When background suppression is used, the labeling efficiency needs to be adjusted based on the number of background suppression pulses |
19 |
100 |
9.10 |
Regions of interest selection should be performed manually as the default approach. Semi-automatic methods may be used if local expertise is available (e.g. using T1 maps) but require extensive validation |
0 |
100 |
9.11 |
Region of interest selection should be performed based on the ASL M0 image or a separately acquired structural dataset |
6 |
93 |
10. Data analysis/reporting |
10.1 |
Cortical renal blood flow values (not whole-kidney) should be reported, separately for left and right kidney |
0 |
100 |
10.2 |
Medullary renal blood flow values are not considered reliable with current measurement approaches |
14 |
89 |