Table 5.

Camera calibration factor and dead-time constant (± SE) derived from segmented SPECT images

Phantom 18–680 keV wide spectrum & TEW
Region-of-interest segmentation Threshold segmentation
Calibration factor (cps/MBq) Dead-time constant (μs) Calibration factor (cps/MBq) Dead-time constant (μs)
Air-3D 9.51 ± 0.06 0.570 ± 0.014 9.25 ± 0.05 0.577 ± 0.012
CTDI-10 9.58 ± 0.04 0.535 ± 0.010 9.30 ± 0.07 0.578 ± 0.016
CTDI-16 9.68 ± 0.06 0.522 ± 0.023 9.36 ± 0.08 0.512 ± 0.032
CTDI-32 9.69 ± 0.04 0.538 ± 0.009 9.25 ± 0.05 0.516 ± 0.016
Jaszczak 9.39 ± 0.04 0.568 ± 0.011 9.38 ± 0.05 0.565 ± 0.013

Method B (full-range CF), with wide-spectrum (18–680 keV) acquisition count rate, was used to derive the calibration factor and the dead-time constant. TEW: Triple-energy window scatter correction