Skip to main content
. Author manuscript; available in PMC: 2021 Jul 1.
Published in final edited form as: ASAIO J. 2020 Jul;66(7):818–824. doi: 10.1097/MAT.0000000000001057

Figure 3: Changes in Leucocyte Counts in the 4 test groups.

Figure 3:

The values are expressed as the mean and standard deviation. There was a decline and rebound in all groups except for the Control group in which there was an elevation in counts after 2 hours of ECC. Leucocyte counts returned to baseline levels by 96 hours in all groups. There was a significant difference in leucocyte counts between Control and Air groups*: p=0.0122 at 120 min and p=0.0122 at 6 hours, respectively. There was also a significant difference in leucocyte counts between Air and 500ppm groups¤: p=0.0137 at 120 min and p=0.0081 at 6 hours, respectively. There was no difference between the Control and 500ppm groups.