Table 2:
Patient | Foramen | Guidance | Time (min) | Obliquity in the Sagittal Plane | Technical Successa | Hydrodissection | Maximal Temperature (°C) |
---|---|---|---|---|---|---|---|
1 | T12–L1 | Fluoro | 11 | 40° | Yes | Yes, effective | 39 |
2 | L1–L2 | Fluoro | 11 | 40° | Yes | Yes, effective | 41 |
3 | T12–L1 | CT & fluoro | 9 | 36° | Yes | Yes, effective | 45 |
4 | L1–L2 | Fluoro | 12 | 35° | Yes | No | 45 |
5 | L2–L3 | Fluoro | 10 | 43° | Yes | Yes, effective | 43 |
L3–L4 | Fluoro | 6 | 38° | Yes | No | 44 | |
6 | T12–L1 | CT & fluoro | 7 | 42° | Yes | Yes, effective | 41 |
7 | T12–L1 | Fluoro | 8 | 44° | Yes | Yes, ineffective | 44 |
8 | L1–L2 | Fluoro & CBCT | 38 | 33° | No (4 mm too cranial and 5 mm too lateral) | Yes, ineffective | 39 |
9 | T12–L1 | Fluoro | 5 | 36° | Yes | Yes, effective | 43 |
10 | L3–L4 | Fluoro | 8 | 47° | Yes | Yes, effective | 38 |
11 | L2–L3 | CT & fluoro | 9 | 37° | Yes | Yes, effective | 39 |
12 | T12–L1 | CT & fluoro | 7 | 47° | Yes | Yes, effective | 45 |
13 | L2–L3 | CT & fluoro | 7 | 44° | Yes | No | 42 |
Note:—Fluoro indicates fluoroscopy; CBCT, conbeam CT.
Technical success was defined by a position of the thermosensor in the midline on anteroposterior view and at the midportion of the posterior wall on lateral view.