Skip to main content
. 2017 Jan 9;12(5):478–485. doi: 10.1016/j.ajps.2017.01.001

Table 1.

Inter-run accuracy and precision following a 3-batch validation of JDR and gemcitabine in rat plasma (data were mean ± SD, n = 18).

Analytes Nominal conc. (ng/ml) Intra-day Inter-day
Measured concentration (ng/ml) Precision (RSD%) Accuracy (RE%) Measured concentration (ng/ml) Precision (RSD%) Accuracy (RE%)
JDR
15.00 15.35 ± 1.4 9.4 −1.6 15.12 ± 0.5 3.4 −3.1
125.00 116.13 ± 3.8 3.2 −7.1 112.60 ± 4.2 3.7 −9.9
3200.00 3573.15 ± 61.5 1.7 11.6 3083.75 ± 25.9 0.8 1.8
Gemcitabine
12.00 11.26 ± 1.2 10.4 4.3 11.02 ± 0.3 2.3 2.0
125.00 124.66 ± 3.9 3.1 −0.3 129.96 ± 6.9 5.3 4.0
3200.00 2912.59 ± 148.1 5.1 −9.0 2931.04 ± 26.2 0.9 −8.4