Skip to main content
. 2020 Feb 11;2020:1684172. doi: 10.1155/2020/1684172

Table 3.

Figures of merit of the MCR-ALS regression model for the validation set of CHZ, CRZ, HAL, IMP, and PTN.

Parameters HAL IMP CHZ CRZ PNT
Accuracy (mean ± SD)a 99.3 ± 1.45 99.8 ± 1.55 99.8 ± 1.13 100.1 ± 1.45 99.9 ± 0.65
Precision repeatability (RSD%)b 1.09 1.32 0.89 0.67 0.78
Intermediate precision (RSD%)c 1.46 1.56 1.13 1.44 0.65
RMSEP 6.56 × 10−2 6.17 × 10−2 1.65 × 10−2 7.81 × 10−2 5.54 × 10−2
SEP 6.34 × 10−2 5.96 × 10−2 1.59 × 10−2 7.54 × 10−2 5.35 × 10−2
Bias 6.96 × 10−3 −1.74 × 10−2 −2.36 × 10−3 2.07 × 10−2 6.74 × 10−3
RE (%) 1.33 1.35 0.75 1.42 0.67
Correlation coefficient (r2) 0.9994 0.9995 0.9997 0.9996 0.9998

aThe mean and standard deviation for 15 determinations. bThe intraday relative standard deviation (n = 3), an average of three different concentration repeated three times within the same day. cThe interday relative standard deviation (n = 3), an average of three different concentration repeated three times in three different days. Low concentrations: 1 μg/mL for HAL, IMP, and CRZ, 0.5 μg/mL for CHZ, and 5 μg/mL for PTN. Intermediate concentrations: 5 μg/mL for HAL, IMP, and CRZ, 2.5 μg/mL for CHZ, and 10 μg/mL for PTN. High concentration: 10 μg/mL for HAL, IMP, and CRZ, 5 μg/mL for CHZ, and 20 μg/mL for PTN.