Table 3.

Accuracy and precision of SH-1242 determinations in rat and mouse quality control (QC) samples.

Batch Rat Plasma Mouse Plasma
Theoretical Concentration (ng/mL) Theoretical Concentration (ng/mL)
LLOQ LQC MQC HQC HQC a LLOQ LQC MQC HQC HQC a
1 2 40 800 800 2 4 40 800 800
Intra-day (n = 6)
Mean 1 1.98 41.9 759 764 1.95 4.06 39.7 752 757
Precision (CV%) b 12.7 4.64 4.41 3.85 1.39 2.56 2.55 2.96 3.09 4.47
Accuracy (RE%) c 0.38 −0.83 4.83 −5.15 −4.48 −2.42 1.42 −0.83 −6.02 −5.35
Inter-day (n = 30)
Mean 1.04 2.04 41.9 777 743 1.99 4.04 40.7 780 782
Precision (CV%) b 12.8 5.04 5.03 4.24 2.84 5.87 4.43 3.57 5.27 6.72
Accuracy (RE%) c 3.95 2.22 4.63 −2.83 −7.15 −0.73 0.91 1.63 −2.53 −2.30

a Analyzed after a ten-fold dilution with blank plasma. b CV(%) = (standard deviation/mean) × 100. c RE(%) = [(calculated concentration – theoretical concentration)/theoretical concentration] × 100. RE: relative error, LQC and MQC: low and mid QC.