Table 6.

Stability of SH-1242 QC samples under typical storage conditions.

Batch Rat Plasma Mouse Plasma
Theoretical Concentration (ng/mL) Theoretical Concentration (ng/mL)
LQC MQC HQC LQC MQC HQC
2 40 800 4 40 800
Benchtop stability at room temperature (25 °C) for 24 h (n = 3)
Mean 1.87 37.3 737 4.21 41.7 810
Precision (CV%) a 6.95 4.35 2.62 9.47 4.38 5.57
Accuracy (RE%) b –6.33 –6.83 –7.88 5.25 4.25 1.25
Autosampler stability at 4 °C for 3 days (n = 3)
Mean 2.00 38.9 773 4.19 42.6 863
Precision (CV%) 8.27 6.05 3.16 4.90 5.15 0.699
Accuracy (RE%) –0.167 –2.75 –3.33 4.67 6.50 7.83
Freeze-thaw stability (3 cycles, n = 3)
Mean 1.84 41.97 780 4.45 42.4 840
Precision (CV%) 3.85 4.20 3.57 2.00 4.32 1.33
Accuracy (RE%) –7.83 4.92 –2.50 11.3 6.00 5.00
Long term stability at 4 °C for 2 weeks (n = 3)
Mean 2.02 37.3 730 3.84 44.8 865
Precision (CV%) 8.54 2.56 0.825 12.0 1.77 4.37
Accuracy (RE%) 0.833 –6.75 –8.71 –4.08 12.0 8.08

a CV(%) = (standard deviation/mean) × 100. b RE(%) = [(calculated concentration – theoretical concentration)/theoretical concentration] × 100.