Skip to main content
. 2019 Nov 28;10(1):49–59. doi: 10.1016/j.jpha.2019.11.006

Table 2.

Results of the experiment performed to determine the LOQ of the GC-MS method for acetazolamide (AZM) in human urine.

Added AZM (μM) PAR 581/584 Measured AZM (μM) Precision (RSD, %) Recovery (%) PAR 83/86 Measured AZM (μM) Precision (RSD, %) Recovery (%)
0 0.006 ± 0.004 1.57 ± 1.05 66.9 not applicable 0.008 ± 0.002 2.26 ± 0.57 25.2 not applicable
25 0.121 ± 0.006 31.7 ± 1.57 5.0 120.5 0.114 ± 0.008 32.3 ± 2.26 7.0 120
50 0.217 ± 0.007 56.9 ± 1.83 3.2 110.7 0.200 ± 0.004 56.6 ± 1.13 2.0 108.7
100 0.396 ± 0.010 103.8 ± 2.62 2.5 102.2 0.367 ± 0.019 103.9 ± 5.4 5.2 101.6
150 0.593 ± 0.011 155.4 ± 2.88 1.9 102.6 0.550 ± 0.015 155.7 ± 4.2 2.7 102.3
200 0.777 ± 0.019 203.6 ± 5.0 2.5 101.0 0.723 ± 0.047 204.6 ± 13.3 6.5 101.2