. 2020 Jan 15;8(1):e001067. doi: 10.1136/bmjdrc-2019-001067

Table 2.

Statistical results for the investigated test strip lot compared against the manufacturer’s reference measurement procedure: percentage of results within ISO 15197 accuracy criteria, minimal deviation from the reference measurement procedure’s results containing at least 95% of values and relative bias according to Bland and Altman (systems are ranked alphabetically)

# BGMS Reference measurement procedure ±15 mg/dL/±15% ±10 mg/dL/±10% ±5 mg/dL/±5% Minimal deviation containing at least 95% of values Relative bias N
% % % mg/dL/% %
a ABRA GOD 91.0 71.0 36.5 ±16.8 −2.9 200
b Accu-Chek Guide HK 100.0 97.5 82.0 ±8.4 −1.6 200
c AURUM GOD 97.5 88.0 55.5 ±13.5 2.6 200
d CareSens Dual GOD 96.5 84.5 48.0 ±13.8 0.8 200
e CERA-CHEK 1CODE GOD 94.5 82.0 51.0 ±15.2 −1.3 200
f ContourNext One GOD 100.0 99.5 77.0 ±7.7 3.3 200
g eBsensor GOD 89.0 73.5 45.5 ±19.7 3.9 200
h FreeStyle Freedom Lite GOD 99.5 91.5 46.5 ±11.1 −6.6 200
i GL50 evo GOD 97.5 86.5 59.5 ±13.4 1.7 200
j GlucoCheck GOLD GOD 100.0 92.0 54.5 ±11.2 −4.5 200
k GlucoMen areo 2K GOD 98.0 86.0 49.0 ±13.3 5.7 200
l GluNEO HK* 93.0 77.5 35.0 ±16.2 −0.6 200
m MyStar DoseCoach GOD 95.0 85.0 56.0 ±14.4 4.3 200
n OneTouch Verio Flex GOD 99.0 93.5 58.5 ±11.9 4.2 200
o Pic Gluco Test GOD 98.0 87.5 54.5 ±12.2 3.3 200
p Rightest GM700S GOD 99.5 96.0 67.0 ±9.4 2.1 200
q TRUEyou GOD 99.0 85.5 56.0 ±13.6 0.9 200
r WaveSense JAZZ Wireless GOD 95.0 84.0 55.0 ±14.4 4.2 200

*No information about the manufacturer’s reference measurement procedure was available at the time of manuscript submission. Based on literature research19 and the investigator’s experience regarding reliability of measurement results, the HK-based procedure was assigned as primary reference measurement procedure for system accuracy evaluation.

GOD, glucose oxidase; HK, hexokinase.