Mean settling values after 700 N static loading in each group and multiple comparisons using Tukey’s honestly significant difference (HSD).

Group Settling Values
Mean ± SD (μm)
Group Comparisons †
Ext 0.8 ± 0.45


Ext < Int-H1, Int-H2 < Int-O2 < Int-O1


Settling value
= (total lengths of the implant-abutment
assemblies at 30 Ncm-③)
minus
(total lengths of the implant-abutment
assemblies after 700 N static loading)
Int-H2 11.2 ± 0.84
Int-H1 10.2 ± 0.84
Int-O2 19.2 ± 4.21
Int-O1 25.6 ± 2.97