Table 2. Intra- and inter-day (n=5) accuracy and precision of QC samples of sumatriptan in plasma.

Nominal Concentration (ng/mL) Accuracy (%) Precision (RSD, %)
Intra-day Inter-day Intra-day Inter-day
0.3 (LLOQ) 95.4 99.5 7.53 11.2
0.9 (LQC) 101.1 103.2 4.74 4.99
30 (MQC) 102.0 102.5 2.39 2.40
80 (HQC) 97.0 93.7 5.33 7.64

LQC, low quality control; MQC, middle quality control; HQC, high quality control.