Skip to main content
. 2020 Feb 20;11:90. doi: 10.3389/fphys.2020.00090

TABLE 3.

Results of the validation experiment during normal walking.

Parameter* Values
Absolute Error
Relative Error
IMU Vicon Mean ± STD Mean ± STD
SL (m) 1.37 ± 0.14 1.33 ± 0.14 0.02 ± 0.03 1.6 ± 2.1%
FCmax(cm) 11.7 ± 1.2 12.4 ± 1.7 −0.7 ± 1.4 −5.6 ± 11.2%
VGait (m/s) 1.17 ± 0.22 1.19 ± 0.24 −0.01 ± 0.02 −0.8 ± 1.6%
Θ (°) 9.3 ± 2.6 9.5 ± 2.8 −0.2 ± 3.3 −1.9 ± 34.9%
SW (cm) 16.5 ± 4.7 7.6 ± 2.7 9.1 ± 4.4 118.4 ± 57.8%
nStepsTurning** (−) 7.2 ± 2.6 5.5 ± 3.0 1.7 ± 0.6 30.9 ± 10.9%
Tstance*** (s) 0.69 ± 0.10 0.72 ± 0.09 −0.02 ± 0.03 −2.9 ± 4.5%
Tswing*** (s) 0.46 ± 0.04 0.44 ± 0.03 0.02 ± 0.04 4.4 ± 8.5%
TDL*** (s) 0.24 ± 0.10 0.16 ± 0.04 0.09 ± 0.07 56.5 ± 43.3%
ncycle (spm) 105.3 ± 9.9 105.5 ± 8.6 −0.9 ± 4.5 −0.9 ± 4.3%
Tcycle (s) 1.15 ± 0.12 1.16 ± 0.11 0.00 ± 0.03 −0.1 ± 2.9%
distarm (m) 0.66 ± 0.19 0.67 ± 0.22 −0.01 ± 0.11 −0.8 ± 16.8%

The IMU sensors and Vicon markers were all attached to the body and the gait parameters were calculated independently with both systems. The absolute error ± standard deviation (STD) is presented as the difference between IMU and Vicon. Abbreviations are listed in Table 2. *RStanceToSwing, dev{Tcycle} and Aswing,arm need no validation, directly calculated from validated parameters/sensor readings. **Visual inspection as reference. ***Period T is validated instead of P = T/Tcycle (% of gait cycle).