Skip to main content
. 2020 Feb 25;14:783–793. doi: 10.2147/DDDT.S239458

Table 2.

Inter-Day and Intra-Day (Precision and Accuracy) of the Developed Methodology

HLM Matrix 5 ng/mL (LLQC) 15 ng/mL (LQC) 150 ng/mL (MQC) 400 ng/mL (HQC)
Inter-Day Assaya Intra-Day Assayb Inter-Day Assay Intra-Day Assay Inter-Day Assay Intra-Day Assay Inter-Day Assay Intra-Day Assay
Mean 4.99 4.98 14.85 14.75 151.6 152.20 400.95 402.21
SD 0.23 0.15 0.37 0.27 1.90 2.02 6.12 2.71
Precision (% RSD) 4.70 3.04 2.49 1.82 1.26 1.33 1.53 0.67
% Accuracy −0.18 −0.38 −0.98 −1.66 1.07 1.47 0.24 0.55

Notes: aAverage of six repeats for 3 days. bAverage of 12 repeats on the same day.