Skip to main content
. 2020 Feb 18;2020:5382739. doi: 10.1155/2020/5382739

Table 3.

Intubation details in scenario C pharyngeal obstruction. Data are presented as median (IQR) or as number (%).

Parameter A B C p values for between-device differences p
Macintosh McGrath TruView PCD A versus B A versus C B versus C
Success rate of intubation attempt (%) 69 (74%) 90 (97%) 67 (72%) 0.023 1.0 0.011 <0.001

Duration of 1st intubation attempt (s) 21 (19–29) 18 (18–25) 30 (23–39) 0.003 0.001 <0.001 <0.001

Cormack-Lehane grade 82 (88%) 93 (100%) 92 (99%) 1.0 1.0 1.0 1.0
1 11 (12%) 0 (0%) 1 (1%)
2 0 (0%) 0 (0%) 0 (0%)
3 0 (0%) 0 (0%) 0 (0%)
4 1 (1–1) 1 (1–1) 1 (1–1)
Median (IQR)

POGO score 80 (78–100) 100 (95–100) 100 (84–100) <0.001 <0.001 0.817 <0.001

Number of optimization manoeuvres (%) 73 (79%) 92 (99%) 39 (42%) 0.021 <0.001 <0.001 <0.001
0 6 (7%) 1 (1%) 46 (50%)
1 14 (15.0%) 0 (0%) 8 (9%)
2 0 (0–0) 0 (0–0) 1 (0–1)
Median (IQR)

Severity of dental compression (%) 21 (23%) 37 (40%) 2 (2%) 0.006 <0.001 <0.001 <0.001
0 53 (57%) 53 (57%) 41 (44%)
1 19 (21%) 3 (3%) 50 (54%)
2 1 (1–1) 1 (0–1) 2 (1–2)
Median (IQR)

Ease of use (1–100) 27 (14–38) 20 (11–24) 39 (24–49) <0.001 <0.001 <0.001 <0.001