Table 2:

Sample acute stroke protocol for 64-section MDCT scanner: NCCT, vertex-to-arch CTA without cardiac component, and shuttle-mode CTP

Location Section Thickness (mm) Image Spacing (mm) Pitch kV mA Gantry Rotation Time (s) Contrast Administration
NCCT C1 to vertex 5 5 0.531:1 120 250 0.7
CTA Vertex to 1 cm below carina 1.25 0.625 0.516:1 120 200–660 0.5 4 mL/s; volume, 60-mL contrast