TABLE 1.

Intra-day (n = 6) and inter-day (n = 18) precision and accuracy for QC samples of GDC-0084 in human plasma and CSF

Intra-day
Inter-day
Nominal (ng/mL) Mean Concentration (ng/mL) %CV %Accuracy Mean Concentration (ng/mL) %CV %Accuracy
Human Plasma 0.2 0.208 1.4 104.0 0.211 4.2 105.5
0.5 0.520 1.4 104.0 0.530 3.2 105.8
10 10.3 3.2 103.0 10.7 4.6 107.0
160 157 1.8 98.1 161 3.9 100.6

Human CSF 0.2 0.209 4.0 104.5 0.189 9.3 94.5
0.5 0.494 1.7 98.8 0.469 4.7 93.8
10 9.84 1.4 98.4 9.70 2.1 97.0
160 156 1.2 97.5 154 2.0 96.3

CV, coefficient of variation.