TABLE II. Insertion Loss Through One Window of the CLINIcell Chamber at a 45° Angle of Insonation, Across 1–7-MHz Transmit Frequencies. Values are Mean (Standard Deviation) of Four Measurements.
| Frequency (MHz) | Insertion loss (dB) |
|---|---|
| 1.0 | 0.92 (0.05) |
| 1.5 | 2.04 (0.06) |
| 2.0 | 4.36 (0.08) |
| 2.5 | 10.45 (0.28) |
| 3.0 | 14.30 (0.39) |
| 3.5 | 4.34 (0.11) |
| 4.0 | 2.63 (0.05) |
| 4.5 | 2.72 (0.09) |
| 5.0 | 3.22 (0.17) |
| 5.5 | 3.64 (0.14) |
| 6.0 | 4.25 (0.03) |
| 6.5 | 7.23 (0.16) |
| 7.0 | 10.50 (0.57) |