Table 3. Cook–Swartz Doppler probe diagnostic test evaluation.
Statistic | Value | 95% CI |
---|---|---|
False-positive rate | 7% | 1–24% |
False-negative rate | 25% | 3–65% |
Sensitivity | 75% | 35–97% |
Specificity | 93% | 76–99% |
Positive likelihood ratio | 10.12 | 2.52–40.75 |
Negative likelihood ratio | 0.27 | 0.08–0.90 |
Abbreviation: CI, confidence interval.
Note : Calculated statistical values according to Cook–Swartz Doppler probe diagnostic test results and the pedicle blood flow status. False-positive and false-negative rates, sensitivity, specificity, and positive and negative likelihood ratios were assessed.