Skip to main content
. Author manuscript; available in PMC: 2020 Mar 9.
Published in final edited form as: Faraday Discuss. 2017 Aug 24;200:579–598. doi: 10.1039/c7fd00021a

Figure 2.

Figure 2.

Comparison of measurements by particle-phase instruments (a) FIGAERO I CIMS mass vs SMPS mass (assumed density = 1.4 g cm−3), shaded region is uncertainty. (b) Time-series of elemental ratios FIGAERO I CIMS to TD-AMS. (c) Histograms of volatility distributions, directly measured by FIGAERO I- CIMS (black solid line) and TD-AMS (gray dashed line), and estimated from FIGAERO I CIMS ions using Eq. 3 as solid green line.