Skip to main content
. 2020 Feb;10(2):314–325. doi: 10.21037/qims.2019.12.11

Table 5. Reproducibility of SPECTRA consensus method when measuring with XtremeCT I and XtremeCT II for Dataset 3 containing n=10 joints.

Measure ICC Lower 95% Upper 95% PE (SD) PE (%CV)
JSV (µm3) 0.85 0.407 0.963 8.06 8.03
JSW (mm) 0.959 0.833 0.99 0.08 3.97
Min (mm) 0.937 0.757 0.984 0.13 13.2
Max (mm) 0.636 -0.332 0.907 0.06 2.19
AS (–) 0.896 0.585 0.974 0.42 12.59

JSV, joint space volume; JSW, joint space width; Min, minimum joint space width; Max, maximum joint space width; AS, joint asymmetry.