Skip to main content
. 2020 Jan;6(Suppl 1):S275–S284. doi: 10.21037/jss.2019.06.09

Figure 2.

Figure 2

Pullout strength of endoscopic PEEK cage tested in tension. Plot shows load (N) versus displacement (mm). The device withstands a tensile load of 665 N (66.5 kg) to failure by pullout due to bone stripping (shown in the left two images). PEEK, polyether ether ketone.