. 2019 Sep 4;75(3):244–252. doi: 10.1136/thoraxjnl-2019-213303

Table 4.

Ventilatory settings

Group Home, N=25 Hospital, N=28
Baseline 3 months 6 months Baseline 3 months 6 months
IPAP, cm H2O 21.0±2.8 22.1±2.9* 23.6±2.3* 24.3±3.6† 24.7±3.3† 25.7±3.4*†
EPAP, cm H2O 4.5±0.8 4.6±0.9 4.6±0.9 5.7±1.2† 5.8±1.2† 6.0±1.3†
IPAP–EPAP, cm H2O 16.5±2.6 17.5±2.5* 19.0±2.1* 18.6±3.3† 18.9±2.8† 19.7±2.7*
BURR, breaths/min 13.5±2.5 13.8±2.2 13.9±2.0 15.6±2.9† 15.3±2.9† 15.4±3.0†

*Significant increase from baseline to 3 months or from 3 months to 6 months.

†Significant difference between the groups at equal time points.

BURR, backup respiratory rate; EPAP, expiratory positive airway pressure;IPAP, inspiratory positive airway pressure.