Table 1: Diagnostic performance of dynamic MPICT.

Summary of studies comparing dynamic stress CT myocardial perfusion imaging (MPI) against invasive coronary angiography (ICA) with fractional flow reserve (FFRcath) or other noninvasive functional tests. Sensitivity (sens), specificity (spec) and area under the curve (AUC) are listed on a per-vessel basis, and compared to CT angiography, if data available. Myocardial blood flow (MBF, ml/min/100cc myocardium), single/dual source computed tomography (SSCT/DSCT), coronary flow rate (CFR).

Author
(year)
N CT system
(vendor)
Dose
(mSv)
MBF
Cut-off
Reference CTA MPICT
Sens
(%)
Spec
(%)
AUC Sens
(%)
Spec
(%)
AUC
Bamberg 2011 33 128-slice DSCT (Siemens) 75 ICA/FFRcath 100 (88-100) 51 (39-63) 93 (77-99) 87 (76-94)
Wang 2012 30 128-slice DSCT (Siemens) 9.5±1.3 NR MPISPECT 90 51 100 75.7
Weiniger 2012 10 128-slice DSCT (Siemens) 12.8±2.4 Visual MPIMRI 86 98
MPISPECT 84 92
Huber 2013 32 256 SDCT (Philips) 9.5 164 ICA >75% FFRcath ≤ 0.75 76 (57-90) 100 (95-100) 0.86
Rossi 2014 99 128-slice DSCT (Siemens) 78 FFRcath ≤ 0.75 61 (77-94) 77 (84-97) 85 (69-94 89 (78-95) 0.95
Kono 2014 42 128-slice DSCT (Siemens) 9.4 0.85 (index) FFRcath ≤ 0.80 98 70 0.85
Kikuchi 2014 32 320-slice SSCT (Canon) 12.8±2.9 2.97 (CFR) MPIPET 86 92 0.90
Tanabe 2016 53 256 MDCT (Philips) 10.5-10.6 92 MPISPECT (N=25) 95 (52-100) 72 (53-91) 0.87
98 MPIMRI (N=28) 78 (67-97) 80 (58-86) 0.89
Coenen 2017 74 128-slice DSCT (Siemens) 9.3-1.8 0.71 (index) ICA/FFRcath ≤0.80 78 (65-90) 49 (38-61) 0.70 73 (61-86) 68 (56-80 0.75
Rossi 2017 115 128/192-DSCT (Siemens) 6.0-10.3 75 (index) ICA/FFRcath ≤0.80 35 (21–53) 95 (81–99) 0.65 89 (76-96) 73 (59-83 0.85
Pontone 2019 85 256-slice SSCT (GE) 5.3±0.7 101 ICA/FFRcath 83 75-91) 66 (9-73) 0.83 73 (63-83) 86 (81-91) 0.88
Nishiyama 2019 38 256-slice SSCT (Philips) 10.2±1.2 126 ICA/FFRcath <0.75 96 (88–100) 57 (46–67) 0.84 83 (68–98) 93 (88–98) 0.96
Alessio 2019 34 256-slice SSCT (GE) 8.4±1.1 126 82-Rubidium PET 75 83 NR