Skip to main content
. Author manuscript; available in PMC: 2020 Mar 12.
Published in final edited form as: Nat Protoc. 2019 Apr 5;14(5):1455–1488. doi: 10.1038/s41596-019-0146-6

Table 1.

Troubleshooting guidance

Step Problem Possible reason Possible solution
Preparation/running
experiments
Worms leaving the wells
  1. Male worm

  2. Hyperactive hermaphrodite

Fill moats around the wells with copper sulfate
Constructing the Blue Light Stimulus for Sleep Perturbation,
12
The power supply is turned on but the LEDs do not turn on
  1. Voltage is set too low

  2. Wired LEDs may be improperly connected

  1. Increase the voltage

  2. Using a multimeter, check the resistance between the positive terminal of the first LED and the negative terminal of the second LED. Properly wired LEDs should have a finite resistance between the input and output terminals.

Setting Imaging System Parameters, 38 The image is too dark
  1. Exposure time is too short

  2. Aperture diaphragm is too small

  3. LEDs not properly positioned and are too far from sample

  4. Voltage of the power supply is too low

  1. Increase the exposure time (Device properties- >Exposure time). Should be less than 1 second, longer exposures result in motion blurring due to animal movement.

  2. Open aperture

  3. Re-position LEDs

  4. Increase the voltage of the power supply

Setting Imaging System Parameters, 38 A white ring is visible at the perimeter of the wells Scattering from the illumination LEDs Ensure the illumination LEDs are not pointing upward but instead are directed horizontally, parallel to the sample plane. The angle between the illumination LED direction and the sample should be minimized to reduce scattering of illumination LEDs onto the camera sensor.