Skip to main content
. 2020 Mar;61(3):461–468. doi: 10.2967/jnumed.119.231308

TABLE 5.

CT Absorbed Doses Determined Using Default Protocols and a Standardized Protocol for Mice and Rats

Default (mGy)
Standard (mGy)
Measured dose difference default to standard (%)
Scanner Mouse Rat Mouse Rat Mouse Rat
1 11±0.10 7±0.10 20±0.09 13±0.16 +77 +86
2 40±0.11 28±0.02 31±0.23 21±0.08 −23 −23
3 59±0.03 48±0.11 39±0.23 28±0.08 −34 −42
4 32±0.18 15±0.10 56±0.76 25±0.05 +71 +60
5 216±0.02 100±0.17 41±0.02 31±0.03 −81 −69

Results expressed as mean ± standard deviation, n = 3.