Table 2.

Calibration curve equations, linear ranges, detection limits, and precision of the MOF-polymer-monolithic-column-based SPME coupled to UHPLC-MS/MS as a detection method for extraction and assessment of SAs from aqueous solution at pH = 5 (n = 5).

Analytes Regression Equation R2 Linear Range
(µg/L)
LOD a
(ng/L)
LOQ a
(ng/L)
RSD b
(%)
SDZ y = 0.2911x − 0.000883 0.9998 0.015–25.0 4.7 14.2 4.7
STZ y = 0.3201x + 0.0243 0.9954 0.015–25.0 3.1 9.35 4.4
SMI y = 0.3595x + 0.0303 0.9989 0.015–25.0 1.7 5.30 6.1
SMZ y = 0.3441x + 0.0207 0.9997 0.015–25.0 2.4 7.42 4.8
SMM y = 0.3448x − 0.00505 0.9994 0.015–25.0 3.5 10.9 4.2
SMX y = 0.3666x + 0.00663 0.9967 0.015–25.0 1.3 4.15 3.7
SIZ y = 0.3617x + 0.0395 0.9972 0.015–25.0 2.3 6.96 2.2
SDM y = 0.3137x + 0.0167 0.9996 0.015–25.0 3.8 11.7 2.1

a Detection of limits and quantification of limits were estimated on the basis of 3:1 and 10:1 signal-to-noise ratios, respectively. b Relative standard deviation (RSD) was monitored using 5.0 µg/L sulfonamides mixed solution.